Appendix 2 — Calibration and Verification Schedule

Parent procedure: 08 – Equipment Maintenance, Calibration and Verification

Purpose

To provide a consolidated schedule of all calibration and verification activities, enabling the laboratory to plan resources, track due dates, and prevent equipment from being used when overdue for calibration.

Instructions

  1. The Technical Manager defines the calibration type, interval, acceptance criteria, and responsible person for each item.
  2. The Quality Manager maintains the schedule, updates dates after each calibration event, and issues reminders at least 30 days before due dates.
  3. This schedule is reviewed at least annually and updated when equipment is added, removed, or when intervals are changed.
  4. Equipment that appears overdue on this schedule must be immediately placed out of service until calibration is completed.

Part A — Calibration and Verification Schedule

Equipment IDDescriptionCalibration / verification typeInternal or externalIntervalAcceptance criteriaResponsibleLast performedNext dueStatusNotes
Analytical instruments
AI-001ICP-OESDaily calibration curveInternalEach runr² ≥ 0.9995; QC recovery ±5%Analyst[Date]Each run☐ Current
AI-001ICP-OESAnnual service and qualificationExternalAnnuallyManufacturer OQ criteriaService engineer[Date][Date]☐ Current · ☐ Due · ☐ Overdue
AI-002HPLCSystem suitability (SST)InternalEach runResolution, tailing, %RSD per SOPAnalyst[Date]Each run☐ Current
AI-002HPLCWavelength accuracy, flow rateInternalAnnuallyPer SOP acceptance criteriaTechnical Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
AI-003Karl FischerCertified water standardInternalEach batchRecovery ±3% of certified valueAnalyst[Date]Each batch☐ Current
AI-004Laser diffractionCertified size standardInternalMonthlyMean within ±3% of certified valueAnalyst[Date][Date]☐ Current · ☐ Due · ☐ Overdue
AI-005BET analyzerCertified BET reference materialInternalQuarterlySA within ±10% of certified value (or per cert.)Analyst[Date][Date]☐ Current · ☐ Due · ☐ Overdue
AI-006XRDPeak position standard (e.g., LaB₆)InternalMonthly2θ deviation ≤ ±0.02°Analyst[Date][Date]☐ Current · ☐ Due · ☐ Overdue
Balances
BA-001Analytical balanceExternal accredited calibrationExternalAnnually≤ ±0.2 mg at test loadsQuality Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
BA-001Analytical balanceInternal verification (ref. weights)InternalMonthly≤ ±0.3 mg at test loadsAnalyst[Date][Date]☐ Current · ☐ Due · ☐ Overdue
BA-001Analytical balanceDaily check (single weight)InternalDaily (before use)≤ ±0.5 mgAnalyst[Date]Daily☐ Current
BA-002Analytical balanceExternal accredited calibrationExternalAnnually≤ ±0.2 mg at test loadsQuality Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
Volumetric delivery
VD-001Pipette 10–100 µLGravimetric calibrationInternalAnnuallyAccuracy ≤ ±1.0%; CV ≤ 0.5%Technical Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
VD-001Pipette 10–100 µLIntermediate gravimetric checkInternalQuarterlyAccuracy ≤ ±1.5% at nominalAnalyst[Date][Date]☐ Current · ☐ Due · ☐ Overdue
VD-002Pipette 100–1000 µLGravimetric calibrationInternalAnnuallyAccuracy ≤ ±0.8%; CV ≤ 0.3%Technical Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
Reference standards
RS-001Reference weight setExternal accredited calibrationExternalEvery 5 yearsPer calibration certificateQuality Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
RS-002Reference thermometerExternal accredited calibrationExternalEvery 2 years≤ ±0.1 °CQuality Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
Environmental monitoring
EM-001Data logger — ICP roomExternal accredited calibrationExternalAnnuallyTemp ≤ ±0.5 °C; RH ≤ ±3%Quality Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
EM-002Data logger — Balance roomExternal accredited calibrationExternalAnnuallyTemp ≤ ±0.5 °C; RH ≤ ±3%Quality Manager[Date][Date]☐ Current · ☐ Due · ☐ Overdue
Supporting equipment
SE-001Water purification systemResistivity checkInternalDaily (before use)≥ 18.2 MΩ·cm (Type I) or per SOPAnalyst[Date]Daily☐ Current
[Add rows]

Part B — Interval Adjustment Record

When calibration intervals are changed from the default, document the justification here.

Equipment IDDescriptionPrevious intervalNew intervalDate changedJustificationApproved by
[ID][Description][e.g., 12 months][e.g., 18 months][Date][e.g., Four consecutive calibrations within tolerance with minimal drift; historical data reviewed][Technical Manager]

Part C — Annual Calibration Calendar

A visual planning aid. Mark the month in which each calibration is due. Helps with resource planning and scheduling external providers.

Equipment IDDescriptionJanFebMarAprMayJunJulAugSepOctNovDec
AI-001ICP-OES annual service
BA-001Balance external cal.
BA-002Balance external cal.
VD-001Pipette calibration
VD-002Pipette calibration
RS-001Weight set cal.
RS-002Ref. thermometer cal.
EM-001Data logger cal.
EM-002Data logger cal.
[Add rows]

● = calibration due in this month. Adjust annually based on actual calibration dates.

Schedule maintained by: [Quality Manager name]Last reviewed: [Date]

:::tip Adapt this template Replace the sample entries with your actual equipment. If you have many items, consider maintaining this schedule electronically (spreadsheet or LIMS) with automated reminders. Regardless of format, a controlled, reviewable version must be available for audit and management review. :::